|
|
|
Applied and Environmental Microbiology, October 2003, p . 6201-6207, Vol . 69, No . 10 Assessment of the Effects of Holding Time and Temperature on Escherichia coli Densities in Surface Water SamplesMisty L . Pope,1* Michelle Bussen,2 Mary Ann Feige,3 Lois Shadix,3 Sharon Gonder,4 Crystal Rodgers,4 Yildiz Chambers,1 Jessica Pulz,1 Ken Miller,1 Kevin Connell,1 and Jon Standridge2 DynCorp, Alexandria, Virginia 22304,1 Wisconsin State Laboratory of Hygiene, Madison, Wisconsin 53718,2 Environmental Protection Agency Technical Support Center, Cincinnati, Ohio 45268-1320,3 Environmental Protection Agency Headquarters, Washington, D.C . 204604 Received 28 February 2003/ Accepted 1 July 2003
Unfortunately, data from evaluations of microbiological indicator density that support current holding time recommendations are limited, particularly for E . coli . In a study comparing fecal coliform levels in samples collected from municipal wastewater and stored for 4 and 24 h at 2 to 4°C (7), results indicated that the means of the 4- and 24-h replicates were within 20% of each other for 24 of 28 sampling events . The researchers considered this range acceptable on the basis of inherent method variability . Researchers evaluating total coliforms in drinking water observed that when the samples were stored at 5°C, densities were 34% lower at 24 than at 0 h (4) . In the same study, when samples were stored at 22°C total coliform densities were 87% lower after 24 h, indicating that both storage time and temperature can impact total coliform densities . In another study evaluating total coliforms in drinking water (8), researchers observed that total coliform densities in some samples stored for 24 h at 22 and 5°C were 47% and 23% lower than densities at 6 h . Total coliform densities were 62 and 33% lower, respectively, after 30 h of storage .
For the USEPA's upcoming Long Term 2 Enhanced Surface Water Treatment Rule (LT2ESWTR), a Federal Advisory Committee recommended that large drinking water treatment systems (serving
To reduce monitoring costs, the Committee recommended that small systems (serving <10,000 people) initially monitor their source water for E . coli in lieu of Cryptosporidium monitoring and only monitor for Cryptosporidium when E . coli levels exceed certain trigger values . Under the Committee's recommended approach, small-system monitoring would begin after completion of the large-system monitoring so that large system monitoring data could be used to further assess the E . coli values that would trigger small system Cryptosporidium monitoring . Large systems can typically analyze E . coli samples within an 8-h holding time either through on-site analyses at the utility's laboratory or by using off-site laboratories within driving distance . However, small systems that do not have onsite E . coli analysis capability and are not within driving distance from a certified laboratory are not able to comply with an 8-h holding time . (The use of delayed incubation procedures, such as Standard Methods 9222E [2], which could increase the allowable sample holding time, may not be appropriate for most small systems, as most such systems do not have personnel trained to perform the sterile-transfer techniques that would be required.) Consequently, the potential widespread violation by small systems of an 8-h E . coli holding time is a significant concern . Analysis of E . coli samples at 24, 30, or 48 h after sample collection might bias results, giving data with values that are lower or higher than those of E . coli analyses performed on the same sample within 8 h of sample collection . To determine whether holding time has an effect on E . coli sample results, studies were conducted in three phases . A summary of each phase is provided in Table 1 . Figure 1 shows the geographical locations of all sites differentiated by phase . For all phases, E . coli density results for each holding time were compared to the initial E . coli density at time 0 and also to the 8-h results . Since both comparisons yielded similar results, only the comparisons to time 0 are presented .
The objective of the phase 2 study was to use real-world sample storage practices (i.e., coolers with wet ice or Utek ice packs) to evaluate additional E . coli samples from surface waters used as drinking water sources . During the phase 2 study, E . coli samples from seven sites within a 2-h driving distance of the Cincinnati laboratory were analyzed at 8, 24, 30, and 48 h after sample collection . The objective of the phase 3 study was to evaluate E . coli holding times from different geographical regions throughout the United States and to evaluate a greater variety of monitoring methods . During the phase 3 study, volunteer utilities organized by the American Water Works Association evaluated E . coli samples from 15 drinking water utility sources at 8, 24, 30, and 48 h after sample collection .
For the sites that serve as drinking water source waters (Lake Winnebago and Rainbow Lake), the utilities provided data from the sample collection points used to collect samples for this study for the following water quality parameters: turbidity, temperature, pH, phosphates, total nitrogen, total organic carbon, and heterotrophic plate counts . When water quality data were not provided by a utility (Lincoln Creek and Wisconsin River), the Wisconsin State Laboratory of Hygiene (WSLH) performed the water quality analyses . WSLH collected water quality samples and E . coli samples at the same time . To determine whether spiking would be necessary, WSLH collected a 1-liter sample 1 day prior to beginning the analysis of samples from each site . To determine ambient E . coli densities, upon collection the WSLH immediately placed each sample on wet ice and transported it to the laboratory for E . coli analysis in triplicate by each method (Colilert and membrane filtration) . On the day the holding time and temperature experiments were started for each site, a sample collection technician drove to the site and collected a 10-liter sample, placed the sample on wet ice, and immediately transported the sample back to the WSLH . For Lake Winnebago and Rainbow Lake, samples of the untreated water were collected directly into sterile, 10-liter cubitainers at the in-plant raw water collection tap . Using a sterile 1-liter bottle multiple times to fill a 10-liter cubitainer, samples were collected at Lincoln Creek and Wisconsin River as composite grab samples . After each sample arrived at WSLH, it was continuously stirred while 100-ml sample aliquots were dispensed into sterile, 160-ml plastic sample containers . Samples for assessment of each holding time, temperature, and method were randomly selected throughout the dispensing process . For waters in which (according to the results of the previous day's analyses) ambient densities of E . coli were below 100 E . coli bacteria per 100 ml, individual 100-ml samples were spiked within 2 h of sample collection . Samples from Lake Winnebago and Rainbow Lake were spiked . Samples from Lincoln Creek and the Wisconsin River had sufficient ambient densities and were not spiked . The following approach was used to spike samples from Lake Winnebago and Rainbow Lake: a single-passage master culture of an environmental isolate of E . coli was verified as E . coli via biochemical reactions, aliquoted to multiple cryovials, and frozen at -70°C . For each site, a frozen vial was thawed, inoculated into Trypticase soy broth, and incubated overnight . Using a spectrophotometer, the overnight culture was diluted to a 0.5 McFarland turbidity standard . From this, two sequential serial 1:1,000 dilutions were made using sterile phosphate buffer . Of the second 1:1,000 dilution, 1 ml was used to spike each 100-ml surface water sample, resulting in a spike of approximately 100 E . coli per 100 ml .
Storage temperatures, holding times, methods, and time 0 sample analysis.
(i) Four sample storage temperatures.
(ii) Four holding times.
(iii) Two methods.
(iv) Time 0 sample analysis.
Phase 2. The utilities provided the temperature and pH data from the sample collection points used to collect samples for this study . The USEPA analyzed the samples for turbidity, phosphates, total nitrogen, and total organic carbon . Water quality samples were collected at the same time as the E . coli samples . Samples were collected and analyzed between March and June of 2002 . To determine whether spiking would be necessary, USEPA personnel collected a 1-liter sample 1 day prior to beginning the analysis of samples from each site . Upon collection, the USEPA personnel immediately placed each sample on wet ice and transported it to the laboratory for E . coli analysis in triplicate by each method (Colilert and mFC/NA-MUG) to measure ambient E . coli concentrations . On the day that the holding time and storage condition experiments were started for each site, utility personnel collected a 10-liter sample and placed the sample on wet ice and the sample was transported back to the USEPA . After the sample arrived at the USEPA, it was spiked (if necessary) and continuously stirred . While being stirred, 100-ml sample aliquots were dispensed into sterile sample containers . Samples for each holding time and method were randomly selected throughout the dispensing process . For waters in which (according to the results of the previous day's analyses) ambient densities of E . coli were below 100 per 100 ml, the 10-liter sample was spiked within 3 h of sample collection . Samples from Bethel, Cincinnati Water Works, Clermont County, Williamsburg, and Williamstown were spiked . Samples from Falmouth and Northern (Newport, Ky.) had sufficient ambient densities and were not spiked . The following approach was used to spike samples from Bethel and Williamstown . A master culture of an environmental isolate of E . coli was verified as E . coli via biochemical reactions and maintained on a nutrient agar slant . For each site, fresh nutrient agar slants were inoculated and incubated for 24 ± 2 h at 35 ± 0.5°C . Dilutions were prepared from the slant with the best growth, and the bulk surface water sample was spiked . The same approach was used to spike samples from Cincinnati Water Works, Clermont County, and Williamsburg, with a laboratory strain of E . coli being used instead of an environmental isolate .
Storage conditions, holding times, methods, and time 0 sample analysis.
(i) Two sample storage conditions (coolants).
(ii) Four holding times.
(iii) Two methods.
(iv) Time 0 samples.
Phase 3. The following flowing stream drinking water sources were evaluated: Potomac River (Fairfax County Water Authority, Herndon, Va.), Fall Creek and White River (Indianapolis Water Company, Indianapolis, Ind.), Mississippi River (Jefferson Parish Water Department, Jefferson, La.), mixed source waters of the Passaic and Ramapo Rivers (Passaic Valley Water Commission, Little Falls, N.J.), Las Vegas Wash (Southern Nevada Water System, Boulder City, Nev.), and Wingra Springs and Willow Creek (Madison, Wis.) . To determine whether spiking would be necessary, the utilities collected an E . coli sample at 24 h prior to sampling for the holding-time study . On the day that the holding-time and storage condition experiments were started, a 10-liter sample was collected, placed on wet ice, and immediately transported back to the utility laboratory . After the sample arrived at the utility, it was spiked (if necessary) and continuously stirred while 100-ml sample aliquots were dispensed into sterile, plastic sample containers . Samples for each holding time and method were randomly selected throughout the dispensing process . For waters in which (according to the results of the previous day's analyses) ambient densities of E . coli were below 100 per 100 ml, the 2-liter sample was spiked within 2 h of sample collection . When necessary, the USEPA provided E . coli spiking suspensions prepared from an environmental isolate . With the exception of WSLH, spiking suspensions were shipped to the utilities on Tuesday, 25 June 2002, and samples were collected and spiked (if necessary) on Wednesday, 26 June 2002 . Personnel at each utility analyzed E . coli samples in triplicate, stored samples in a cooler using a coolant of their choice, and packed the samples as they would for shipment . Four holding times (8, 24, 30, and 48 h) were evaluated, and holding times began from the time that samples were spiked or dispensed into 100-ml aliquots . For all sites, three time 0 samples were analyzed to determine the initial E . coli density . The initial time 0 samples were not exposed to the holding temperatures, because the samples were analyzed prior to being stored . Personnel at each utility evaluated one or more sites using a method(s) of their choice . Methods used to analyze E . coli samples during the phase 3 study included Colilert Quanti-Tray/2000, membrane filtration onto mFC followed by transfer to NA-MUG, membrane filtration onto mEndo followed by transfer to NA-MUG, and membrane filtration using mTEC (Standard Methods 9213D.3) (2) .
Using Dunnett's test (1), E . coli density results for each holding time and temperature combination were compared to the initial E . coli density at time 0 and the 8-h E . coli density results for each method (alpha = 0.05) . Dunnett's test is a statistical method of testing for pairwise differences in means in which control group results are compared to the results for each treatment combination . These pairwise comparisons are run after the GLM because even though GLMs determine whether the concentration of at least one level of a given predictor variable differs significantly from the concentration of at least one other level, GLMs do not specify whether these pairwise differences include the control group . By comparing treatment groups to the control group only, Dunnett's test holds the overall type I error rate to 5% and is more powerful than other tests, such as Tukey or Scheffe, for pairwise comparisons . In this study, the controls (represented by the time 0 results for each method and site) were compared to each treatment combination (i.e., results from a specific holding time and temperature for each method and site) . It was necessary to compare the control to each time and temperature combination, because the time 0 samples did not have an associated holding temperature . A summary of the phase 1 Dunnett's test results for the time 0 comparison is provided in Table 3 .
(ii) Colilert samples stored at 4°C or 10°C.
(iii) Membrane filtration (mFC/NA-MUG) samples stored at 10°C.
(iv) Membrane filtration (mFC/NA-MUG) samples stored at 4°C. The phase 1 results suggest that E . coli samples can be analyzed beyond 8 h after sample collection using some methods (with chilling) while still generating reliable E . coli data . However, a wider variety of surface waters from different regions throughout the United States still needed to be evaluated to verify these preliminary observations . Phases 2 and 3 were conducted to address these issues .
Phase 2. Mean E . coli densities (E . coli per 100 ml) for site, method, holding time, and coolant (wet ice or Utek ice packs) were compared using GLM at an alpha = 0.05 level of significance . Interactions with methods and sites were observed, and as a result, all subsequent statistical analyses were stratified by method and site . Using Dunnett's test, the E . coli density results for each holding-time and coolant (wet ice or Utek ice packs) combination were compared to the initial E . coli density results at time 0 and to the 8-h E . coli density results for each method (alpha = 0.05) . A summary of the phase 2 Dunnett's test results for the time 0 comparison is provided in Table 4 .
Phase 3.
Conclusions. Although not one of the objectives of this study, it was observed through continuous sample temperature monitoring that water samples for E . coli sometimes froze during storage . Because cell lysis may occur under these conditions, precautions should be taken to prevent samples from freezing during storage .
We greatly appreciate the recommendations of Steve Via (American Water Works Association), Stig Regli (USEPA), and Nancy Hall (University of Iowa Hygienic Laboratory) during development of the study design . IDEXX is gratefully acknowledged for providing Colilert and Quanti-Tray/2000 for all analyses conducted during the phase 2 study at USEPA . We greatly appreciate Jennifer Birkenhauer, Jason Burlage, and Richard Hamblin (Oak Ridge Institute for Science and Education), Bonnie Newport and Keith Kelty (USEPA), and Jessica Vande Burgt (City of Milwaukee) for their assistance in collecting and analyzing samples . We appreciate the USEPA's Office of Research and Development for the chemical analysis of source waters from phase 2 . We appreciate the efforts of Ben Meadows, Kyle May, Rashmi Ghei, Jenifer Grabski, and Garen Lott (DynCorp) for their assistance with data review, validation, and analysis, figure development, and manuscript review . We also appreciate the following for participating in phase 3: Richard Talley and Saeed Samavat, Fort Worth Water Department; Philip Johnson and Gerri Murray, City of Phoenix; Yone Akagi, Portland Water Bureau; Peggy Roefer, Maria Johnson, Darlene Jones, and Debbie Schoolcraft, Southern Nevada Water System; Connie Schreppel, Mohawk Valley Water Authority; Sherry Williams, Newport News Waterworks; Meg Carlson and Melissa Billman, Fairfax County Water Authority; Kim Brokering, Indianapolis Water Company; Wayne Koffskey and Bob Grant, Jefferson Parish Water Department; and Linda Tatro, Passaic Valley Water Commission .
What Is Molecular Microbiology?,
What Is Salmonella?,
What Is Bioassay?,
What Is Biofilm?,
What Is Biofilter?,
r,
Bacteria,
i,
Bacterium,
s,
Microorganism,
o,
Microbe,
n,
Bacteriology,
a,
Wastewater,
o,
Escherichia coli,
i,
Antibiotic resistance,
o,
Bactericidal,
o,
Bactericidal,
c,
Bacteria,
a,
Staphylococcus aureus,
a,
Yeasts,
e,
Minimal inhibiting concentration,
s,
Clostridia,
o,
Candida albicans,
e,
Enterococci,
i,
Rhizobacterium,
c,
Antibiotics,
s,
Antimicrobials,
e,
Typhus,
r,
Culture medium,
c,
Candida albicans,
n,
Escherichia coli,
i,
Vancomycin,
i,
Escherichia coli
|
© 2005
Transgalactic Ltd (manufacturer of Bioscreen C software) |
Privacy Statement | P.O. Box
1393, 00101 Helsinki, Finland,
Last modified: May 25, 2005
| ||||||